SCORES:Clinical and angiographic efficacy of a self-expanding nitinol stent in saphenous vein graft atherosclerotic disease: the Stent Comparative Restenosis (SCORES) Saphenous Vein Graft Registry.

作者:国际循环网   日期:2004/8/24 0:00:00

国际循环网版权所有,谢绝任何形式转载,侵犯版权者必予法律追究。

Am Heart J. 2003 May;145(5):868-74. Clinical and angiographic efficacy of a self-expanding nitinol stent in saphenous vein graft atherosclerotic disease: the Stent Comparative Restenosis (SCORES) Saphenous Vein Graft Registry. Kandzari DE, Goldberg S, Schwartz RS, Chazin-Caldie M, Sketch MH Jr;SCORES SVG Registry Investigators. Division of Cardiology, Duke University Medical Center, Durham, NC 27710, USA.

BACKGROUND: The Stent Comparative Restenosis (SCORES) Saphenous Vein Graft (SVG) Registry was a multicenter, prospective registry designed to evaluate the safety and efficacy of a self-expanding, nickel-titanium (nitinol) stent for de novo SVG lesions.

METHODS: In all, 159 patients with de novo vein graft lesions > or =2.75 and < or =4.25 mm in diameter and <30 mm in length underwent stenting with the Radius self-expanding stent. The primary end point was target vessel failure (TVF) at 9 months, which was defined as a composite of procedural failure, death, myocardial infarction, or target vessel revascularization.

RESULTS: Procedural success was achieved in 96.8% of patients, and the 30-day incidence of major adverse cardiac events was 2.5%. The binary rate of restenosis at 6 months was 28.6%. By 9 months, the rate of TVF was 24.5%, and the rate of major adverse cardiac events was 23.1%. The 9-month Kaplan-Meier survival rates for freedom from TVF and target lesion revascularization were 76.0% and 87.9%, respectively. No clinical or angiographic characteristic was predictive of restenosis.

CONCLUSIONS: In de novo atherosclerotic SVG disease, the use of a self-expanding, nitinol stent was associated with high initial procedural success and favorable early and intermediate outcomes. Because few studies have examined the influence of stent composition and design in SVG disease, these findings not only show the safety and efficacy of this self-expanding stent in de novo SVG disease, but also merit further comparison with balloon-expandable stents.

版面编辑:国际循环


SCORES

分享到: 更多


设为首页 | 加入收藏 | 关于我们 | 联系方式 | 招贤纳士
声明:国际循环网( www.icirculation.com)对刊载的所有文章、视频、幻灯、音频等资源拥有全部版权。未经本站许可,不得转载。
京ICP备15014970号-5  互联网药品信息服务资格证书编号(京)-非经营性-2017-0063  京公网安备 11010502033353号  增值电信业务经营许可证:京ICP证150541号
国际循环 版权所有   © 2004-2024 www.icirculation.com All Rights Reserved
公司名称:北京美赞广告有限公司 公司地址:北京市朝阳区朝阳门北大街乙12号天辰大厦1座1409 电话:010-51295530